TKSAGE-600 VHP Generator | ||||
Sterilization time | 120-300min | |||
Sterilization space (m3) | 100-600m3 | |||
External dimensions (w*h*d)/mm) | 450mm*800mm*1000mm | |||
Hydrogen peroxide vaporization rate | 5-30ml/min | |||
Weight (kg) | 60 kg | |||
Max power (W) | 3600W | |||
Fan blast volume (m3/h) | 1500CMH | |||
Air pressure (Pa) | 500Pa | |||
Power supply | 220V,50Hz |
Sterilization principle and features
The TKSAGE-600 hydrogen peroxide sterilizer adopts unique low-temperature flash evaporation technology to vaporize 30%-35% hydrogen peroxide and quickly distribute it in the sterilization space to achieve space surface sterilization. The whole equip - ment is made of stainless steel, with smooth appearance and no dead ends, in line with GMP requirements. The user-friendly opera - tion interface is simple and easy to operate, so users can quickly master the operating skills. The sterilization program adopts fully automatic process control, which can monitor the process temperature, humidity, and hydrogen peroxide concentration in real time and calculate the sterilization effect according to the real time parameters. The equipment can execute 6LOG sterilization program, 12LOG sterilization program, concentration/time program and other options to meet different needs of customers. The sterilization process uses the dew point control method to ensure the vaporization state of the entire sterilization process.
TKSAGE-600 main interface | TKSAGE-600 control parameters |
![]() | ![]() |
VHP sterilization curve
Start: Before sterilization starts, the equipment automatically determines whether the sterilization conditions meet
the sterilization start requirements, i.e.: temperature>15°C, 40%< RH%<60%; Balance: The VHP generator starts
to produce hydrogen peroxide vapor and circulates it through the space until it reaches a preset concentration or
saturation level; Sterilize: The equipment starts to record the sterilization effect until the preset sterilization LOG
value is reached. Degrade: At the end of the sterilization process, the equipment automatically enters the VHP
degradation stage, which starts to catalyze the degradation of the hydrogen peroxide vapor inside the equip-
ment and the cavity until it reaches the set hydrogen peroxide concentration.